Observation |
Suspected Problem |
Possible Normal Pacer Function |
Pacing artifacts delivered
in the middle of normal ORS complexes |
Pacemaker undersensing | Pseudofusion |
Unexplained
pauses after sensed but |
Pacemaker oversensing |
Hysteresis |
Pacer
rate less than lower rate, which occurs at night |
Pacemaker oversensing at night |
Different sleep rate/lower rate |
Rapid pacing rates | Pacemaker oversensing of
EMI in atrial channel or other atria events causing upper rate tracking Pacemaker-mediated tachycardia |
Scanning hysteresis/rate
drop response causing an increased "intervention" pacing rate for vasovagal episodes Activation of rate-response sensors Upper rate tracking of unsuspected atrial tachyarrhythmia or sinus tachycardia Pacer reset due to EMI or low battery Pacer that mode switched due to atria tachyarrhythmia |
DDD pacer operating in VVI mode |
Malfunction of pacer |
Pacer reset due to EMI or
low battery Pacer that mode switched due to atria tachyarrhythmia |
Table 9
Suspected Malfunction Occurring During Normal Pacer Function